Quality assurance of medical accelerators using VMAT technology based on 3D semiconductor detector array ArcCHECK

Journal Of Modern Oncology[ISSN:1672-4992/CN:61-1415/R]

Issue:
2024 02
Page:
308-315
Research Field:
Publishing date:

Info

Title:
Quality assurance of medical accelerators using VMAT technology based on 3D semiconductor detector array ArcCHECK
Author(s):
DENG Jia12LIU Shengyan3HUANG Yun4ZHAO Yaolin2WU Xiangyang1
1.Department of Radiation Oncology,Shaanxi Province Cancer Hospital,Shaanxi Xi'an 710061,China;2.School of Nuclear Science and Technology,Xi'an Jiaotong University,Shaanxi Xi'an 710049,China;3.Yulin Xingyuan Hospital,Shaanxi Yulin 719000,China;4.Xianyang Central Hospital,Shaanxi Xianyang 712000,China.
Keywords:
VMATIMRTMachine QAArcCHECK
PACS:
R730.55
DOI:
10.3969/j.issn.1672-4992.2024.02.020
Abstract:
Objective:To optimize the clinical outcomes of tumor radiotherapy by utilizing the ArcCHECK semiconductor detector to assess the quality control of VMAT technology in medical accelerators.Methods:The ArcCHECK cylindrical semiconductor radiation detector's three-dimensional array was employed with specific dose irradiation techniques.The measured results were used to calculate deviations in gantry isocenter,MLC positioning,and beam flatness symmetry.Dose deviations were correlated with mechanical performance.The experimental results were compared and monitored against routine daily and monthly tests for 12 months.The sensitivity to radiation beam was also evaluated to enhance accelerator radiation monitoring efficiency and accuracy.Results:The gantry isocenter deviation was 0.292 mm±0.016 mm,with an average gantry speed of 3 degrees/second and an average maximum speed deviation of approximately 0.1 degrees/second.The average maximum MLC deviation was 0.392 mm±0.049 mm.The average beam flatness symmetry was 105.55%±0.13%,and the X and Y-axis symmetries were 101.02%±0.21% and 101.11%±0.21%,respectively.The 12-month continuous measurements showed consistent trends with independently verified results.Conclusion:Leveraging the advantages of the ArcCHECK cylindrical semiconductor detector array,it provides a comprehensive measurement approach for VMAT quality control,serving as a means for medical accelerator radiation detection and quality assurance.

References:

[1] PALMA D,VOLLANS E,JAMES K,et al.Volumetric modulated arc therapy for delivery of prostate radiotherapy:comparison with intensity-modulated radiotherapy and three-dimensional conformal radiotherapy[J].International Journal of Radiation Oncology Biology Physics,2008,72(4):996-1001.
[2] KOKSAL M,OZKAN O,HOLDERRIED T,et al.Optimized conformal total body irradiation with VMAT using a linear-accelerator-based radiosurgery treatment system in comparison to the golden standard helical tomotherapy[J].Cancers,2023,15(17):4220.
[3] OSMAN A,MAALEJ NM.Applications of machine and deep learning to patient-specific IMRT/VMAT quality assurance[J].Journal of Applied Clinical Medical Physics,2021,22(8):1-17.
[4] HUNTE SO,CLARK CH,ZYUZIKOV N,et al.Volumetric modulated arc therapy (VMAT):A review of clinical outcomes-What is the clinical evidence for the most effective implementation[J].The British Journal of Radiology,2022,95(1136):20201289.
[5] LIN X,SUN T,LIU X,et al.Comparison of MLC positioning deviations using log files and establishment of specific assessment parameters for different accelerators with IMRT and VMAT[J].Radiation Oncology,2022,17(1):1-11.
[6] ALHARTHI T,ARUMUGAM S,VIAL P,et al.EPID sensitivity to delivery errors for pre-treatment verification of lung SBRT VMAT plans[J].Physica Medica,2019,59:37-46.
[7] MARSAC T,CAMEJO AB,CHIAVASSA S,et al.An effective and optimized patient-specific QA workload reduction for VMAT plans after MLC-modelling optimization[J].Physica Medica,2023,107:102548.
[8] MANS A,SCHURING D,ARENDS MP,et al.The NCS code of practice for the quality assurance and control for volumetric modulated arc therapy[J].Physics in Medicine & Biology,2016,61(19):7221.
[9] MHATRE VR,PATWE PT.Evaluation of ArcCHECK SNC machine QA tool for modern linear accelerator[J].International Journal of Cancer Therapy and Oncology,2017,5:1-10.
[10] THIYAGARAJAN R,NAMBIRAJ A,SINHA SN,et al.Analyzing the performance of ArcCHECK diode array detector for VMAT plan[J].Reports of Practical Oncology and Radiotherapy,2016,21(1):50-56.
[11] CHASWAL V,WELDON M,GUPTA N,et al.Commissioning and comprehensive evaluation of the ArcCHECK cylindrical diode array for VMAT pretreatment delivery QA[J].Journal of Applied Clinical Medical Physics,2014,15(4):212-225.
[12] BARNES M,GREER P.Evaluation of the TrueBeam machine performance check (MPC) beam constancy checks for flattened and flattening filter-free (FFF) photon beams[J].Journal of Applied Clinical Medical Physics,2017,18(1):139-150.
[13] BARNES MP,GREER PB.Evaluation of the truebeam machine performance check (MPC):mechanical and collimation checks[J].Journal of Applied Clinical Medical Physics,2017,18(3):56.
[14] LOW DA.Minimization of target positioning error in accelerator-based radiosurgery[J].Medical Physics,1995,22(4):443-448.
[15] WEILIANG D,JENNIFER W.On the selection of gantry and collimator angles for isocenter localization using Winston-Lutz tests[J].Journal of Applied Clinical Medical Physics,2016,17(1):167-178.
[16] MAMALUI-HUNTER M,LI H,LOW DA.MLC quality assurance using EPID:a fitting technique with subpixel precision[J].Medical Physics,2008,35(6 Part 1):2347-2355.
[17] IBRAHIM S,SESHADRI V,CHARLES P.Assessing the suitability of a two-dimensional array for routine quality assurance checks of flatness and symmetry[J].Physical and Engineering Sciences in Medicine,2020,43(4):1451-1460.
[18] PANT A,MIRI N,BHAGROO S,et al.Monitor unit verification for varian TrueBeam VMAT plans using Monte Carlo calculations and phase space data[J].Journal of Applied Clinical Medical Physics,2023,7(19):e14063.
[19] WOUDSTRA E,HEIJMEN BJM,STORCHI PRM.Automated selection of beam orientations and segmented intensity-modulated radiotherapy (IMRT) for treatment of oesophagus tumors[J].Radiotherapy and Oncology,2005,77(3):254-261.
[20] BREEDVELD S,STORCHI PRM,VOET PWJ,et al.iCycle:Integrated,multicriterial beam angle,and profile optimization for generation of coplanar and noncoplanar IMRT plans[J].Medical Physics,2012,39(2):951-963.
[21] TEOH M,CLARK CH,WOOD K,et al.Volumetric modulated arc therapy:a review of current literature and clinical use in practice[J].The British Journal of Radiology,2011,84(1007):967-996.
[22] PALMA DA,VERBAKEL WFAR,OTTO K,et al.New developments in arc radiation therapy:a review[J].Cancer Treatment Reviews,2010,36(5):393-399.
[23] OTTO K.Volumetric modulated arc therapy:IMRT in a single gantry arc[J].Medical Physics,2008,35(1):310-317.
[24] KUTCHER GJ,COIA L,GILLIN M,et al.Comprehensive QA for radiation oncology:report of AAPM radiation therapy committee task group 40[J].Medical Physics,1994,12:581.
[25] BIGGS PJ,LING CC,PURDY JA,et al.AAPM code of practice for radiotherapy accelerators:report of AAPM radiation therapy task group No.45[J].Medical Physics,1994,21:1093.
[26] CHANG L,HO SY,DU YC,et al.An improved method to accurately calibrate the gantry angle indicators of the radiotherapy linear accelerators[J].Nuclear Inst & Methods in Physics Research A,2007,576(2-3):441-445.
[27] CHANG L,HO SY,WU JM,et al.Technical innovation to calibrate the gantry angle indicators of linear accelerators[J].Journal of Applied Clinical Medical Physics,2001,2(1):54-58.
[28] LING CC,ZHANG P,ARCHAMBAULT Y,et al.Commissioning and quality assurance of rapidarc radiotherapy delivery system[J].Int J Radiat Oncol Biol Phys,2008,72(2):575-581.
[29] MASI L,CASAMASSIMA F,DORO R,et al.Quality assurance of volumetric modulated arc therapy:Evaluation and comparison of different dosimetric systems[J].Medical Physics,2011,38(2):612-621.
[30] KLEIN EE,HANLEY J,BAYOUTH J,et al.Task group 142 report:Quality assurance of medical accelerators[J].Medical Physics,2009,36(9Part1):4197-4212.
[31] LOSASSO T,CHUI CS,LING CC.Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy[J].Medical Physics,1998,25(10):1919-1927.
[32] VT H,KHANNA D,RAJ A,et al.Dosimetric importance of implementing jaw tracking technique in radiotherapy treatment plan execution[J].Asian Pacific Journal of Cancer Prevention,2022,23(4):1397-1403.

Memo

Memo:
陕西省卫生健康科研基金项目(编号:2022D037)
Last Update: 1900-01-01