The 3D printed headrest improves the repeatability of the position of intensity-modulated radiotherapy for nasopharyngeal carcinoma

Journal Of Modern Oncology[ISSN:1672-4992/CN:61-1415/R]

Issue:
2023 08
Page:
1485-1489
Research Field:
Publishing date:

Info

Title:
The 3D printed headrest improves the repeatability of the position of intensity-modulated radiotherapy for nasopharyngeal carcinoma
Author(s):
YAN Yuying1YE Rong1YU Luxin1TANG Xiaoyan1HUANG Fasheng1YANG Chengfan2
1.Cancer Center;2.3D Printing Center,Sichuan Academy of Medical Sciences·Sichuan Provincial People's Hospital,Sichuan Chengdu 610072,China.
Keywords:
3D printed headrestnasopharyngeal carcinomaintensity modulated radiotherapysetup errorresidual translation errorresidual rotation error
PACS:
R739.63
DOI:
10.3969/j.issn.1672-4992.2023.08.018
Abstract:
Objective:To analyze the effect of 3D printed headrest on the reproducibility of intensity modulated radiation therapy (IMRT) for nasopharyngeal carcinoma (NPC).Methods:54 patients with NPC were divided into experimental group (27 cases) and control group (27 cases).The experimental group was fixed with 3D printed headrest combined with head-neck-shoulder thermoplastic membrane,while the control group was fixed with foam soft headrest,carbon fiber pad and head-neck-shoulder thermoplastic membrane.The setup errors in the three directions of anterior-posterior(AP),superior-inferior(SI),left-right(LR)were retrospectively compared between the two groups,and the residual translation errors and residual rotation errors in the 4 regions of interest of basis cranial,cervical 2,cervical 4 and cervical 7 were analyzed by registration.Results:The number of patients needing repeated positioning before the first treatment in the experimental group was significantly less than that in the control group with statistical significance (P<0.05).The setup errors of AP and SI directions in the experimental group were significantly smaller than those in the control group(P<0.05),and there was no significant differencein the LR direction between the two groups (P>0.05).The residual translation errors of cranial base and cervical 2 in the AP,SI and LR directions and cervical 4 and cervical 7 in the AP,LR directions in the experimental group were significantly smaller than those in the control group(P<0.05).There was no significant difference in the residual translation errors in the SI direction of cervical 4 and cervical 7 compared with the control group (P>0.05).The residual rotation errors of cranial base,cervical 2,cervical 4 and cervical 7 in the sagittal,transverse,and coronal sections in the experimental group were significantly lower than those in the control group (P<0.05).In the experimental group,the frequency of absolute value of residual translation error>2 mm in the AP direction of cranial base and the AP,SI and LR directions of cervical 2,cervical 4 and cervical 7 was significantly lower than that in the control group (P<0.05).The frequency of absolute value of residual translation error>2 mm in the SI and LR directions of cranial base was not significantly different from that in the experimental group and in the control group (P>0.05).In the experimental group,the frequency of absolute value of residual rotation errors>2° in the sagittal,transverse,and coronal sections of cranial base,cervical 2,cervical 4 and cervical 7 was significantly lower than that in the control group (P<0.05).Conclusion:The 3D printed headrest combined with head-neck-shoulder membrane fixed position can reduce the setup error,residual translation error and residual rotation error in intensity modulated radiotherapy(IMRT)for nasopharyngeal carcinoma(NPC).it can effectively improve the accuracy of IMRT and decrease the side effects of normal tissue and improve the target therapeutic effect and the quality of life of patients.

References:

[1] 许森奎,姚文燕,严惠莲,等.两种体位固定装置在颈部放疗中的摆位误差比较研究[J].肿瘤预防与治疗,2019,32(6):528-532. XU SK,YAO WY,YAN HL,et al.A comparative study on setting errors of two kinds of position fixing devices in neck radiotherapy [J].Cancer Prevention and Treatment,2019,32(6):528-532.
[2] 丘敏敏,徐圆,钟嘉健,等.鼻咽癌放疗中 T形口腔固定器的研制及其摆位误差和剂量影响研究[J].中国医学物理学杂志,2020,37(11):1367-1372. QIU MM,XU Y,ZHONG JJ,et al.Development of t-shaped oral fixator for radiotherapy of nasopharyngeal carcinoma and study on its setting error and dose effect[J].Chinese Journal of Medical Physics,2020,37(11):1367-1372.
[3] HUANG H,LU H,FENG G,et al.Determining appropriate timing of adaptive radiation therapy for nasopharyngeal carcinoma during intensitymodulated radiation therapy [J].Radiat Oncol,2015,10 (1):192.
[4] 刘晓斌,沈文.影像组学在头颈部肿瘤放疗的研究进展[J].中华放射肿瘤学杂志,2021,30(1):98-101. LIU XB,SHEN W.Research progress of imaging in radiotherapy of head and neck tumors [J].Chinese Journal of Radiation Oncology,2021,30(1):98-101.
[5] WANG H,WANG C,TUNG S,et al.Improved setup and positioning accuracy using a three-point customized cushion/mask/bite-block immobilization system for stereotactic reirradiation of head and neck cancer[J].J Appl Clin Med Phys,2016,17(3):180-189.
[6] 张萍,虞维博,许青,等.探究体质量指数对鼻咽癌患者放疗摆位误差的影响[J].中国癌症杂志,2018,28(5):389-393. ZHANG P,YU WB,XU Q,et al.To investigate the effect of body mass index (BMI) on setting-up error of radiotherapy in patients with nasopharyngeal carcinoma (NPC)[J].Chinese Journal of Cancer,2018,28(5):389-393.
[7] 吴冰,付爽.热塑体膜固定技术在胸腹部肿瘤放疗中的应用[J].中国医学装备,2008,23(12):87-88. WU B,FU S.Application of thermoplastic body membrane fixation technique in radiotherapy of Thoracic and abdominal tumors [J].Chinese Medical Equipment,2008,23(12):87-88.
[8] 薛青,陆维,张艳俊,等.探讨不同固定方式在鼻咽癌放疗中的应用[J].瘤预防与治疗,2019,32(1):67-71. XUE Q,LU W,ZHANG YJ,et al.To investigate the application of different fixation methods in radiotherapy of nasopharyngeal carcinoma (NPC)[J].Tumor Prevention and Treatment,2019,32(1):67-71.
[9] 许森奎,姚文燕,胡江,等.鼻咽癌发泡胶个体化塑形与标准化头枕放疗体位固定精确度比较[J].中华放射肿瘤学杂志,2015,24(2):196-199. XU SK,YAO WY,HU J,et al.Comparison of the accuracy of position fixation between Styrofoam and standard headrest radiotherapy for nasopharyngeal carcinoma[J].Chinese Journal of Radiation Oncology,2015,24(2):196-199.
[10] 招什武,戴振晖,朱远湖,等.鼻咽癌调强放疗两种体位固定方式的摆位误差分析[J].临床医学工程,2018,25(5):575-577. ZHAO SW,DAI ZH,ZHU YH,et al.Setup error analysis of two different postural fixations in intensity modulated radiation therapy for nasopharyngeal carcinoma[J].Clinical Medical Engineering,2018,25(5):575-577.
[11] 傅万凯,刘利彬,陈彩霞,等.发泡胶枕在鼻咽癌调强放疗中的应用[J].现代肿瘤医学,2020,28(19):3413-3416. FU WK,LIU LB,CHEN CX,et al.Application of Styrofoam pillow in intensity modulated radiation therapy for nasopharyngeal carcinoma[J].Modern Oncology,2020,28(19):3413-3416.
[12] 段巧.发泡胶用于头颈部肿瘤头颈肩体位固定放疗的临床效果[J].医疗装备,2018,31(8):81-89. DUAN Q.Clinical effect of Styrofoam on head-neck and shoulder fixed radiotherapy for head-neck tumors [J].Medical Equipment,2018,31(8):81-89.
[13] 刘礼东,杨振,钟美佐,等.肿瘤放疗中塑形垫结合热塑网膜体位固定方式的摆位误差研究[J].中国肿瘤临床,2020,47(4):198-201. LIU LD,YANG Z,ZHONG MZ,et al.Study on position error of plastic pad combined with thermoplastic mesh in tumor radiotherapy[J].Chinese Clinical Oncology,2020,47(4):198-201.

Memo

Memo:
-
Last Update: 1900-01-01