The application of CT simulator and simulator in IMRT position verification of head-and-neck cancers

Journal Of Modern Oncology[ISSN:1672-4992/CN:61-1415/R]

Issue:
2017 06
Page:
948-951
Research Field:
Publishing date:

Info

Title:
The application of CT simulator and simulator in IMRT position verification of head-and-neck cancers
Author(s):
Xiao YouliXie GuanhuiOu YiminNi QianxiZhang Jiutang
Hunan Cancer Hospital/The Affiliated Cancer Hospital of Xiangya School of Medicine,Central South University,Hunan Changsha 410013,China.
Keywords:
CT simulation machinesimulatorIMRTposition verifieserror margin
PACS:
R739.91
DOI:
10.3969/j.issn.1672-4992.2017.06.031
Abstract:
Objective:To explore the method and advantages of the GE LightSpeed RT CT and the Nucletron SIMULIX HQ simulator model in the treatment of head and neck cancer IMRT target region.Methods:Sixty patients fixed by head-neck-shoulder thermoplastic membrane with head and neck tumor IMRT were randomly divided into two groups,using the DRR Frontal&Side image in 30 cases,and 30 cases were treated with the center of the tomograph image in the CT simulator.The position difference between the directions of X (left-light),Y (head-feet),Z (front-back) three directions and the location of the treatment center were measured.The error was Ey,Ez and Ex in three directions.Results:The errors are divided into three grades Ea≤±1mm,±1mm<Eb≤±2mm,Ec>±2mm.CT simulation machine calibration center in the X direction of the error range of the number of cases in turn 8(26.7%),18(60.0%),4(13.3%);Y direction 19(63.3%),8(26.7%),3(10.0%);Z direction 9(30.0%),14(46.7%),7(23.3%).In Simulation calibration center X direction 6(20.0%),19(63.3%),5(16.7%);Y direction 9(30.0%),18(60.0%),3(10.0%);Z direction 8(26.7%),14(46.7%),8(26.7%).The average error of the calibration center of the CT simulator in three directions were 1.39mm 0.53mm 1.48mm Simulation were 1.51mm,1.43mm,1.66mm.60 patients were treated with the center for the first time,and the error of the IGRT was controlled by the 2mm test.Conclusion:IMRT patients with head and neck tumor fixed by head-neck-shoulder thermoplastic membrane,whether it is to take a CT simulation or a simulation of the positioning machine reset,the overall highest in the three-dimensional direction is only 40% of the displacement accuracy of the patients in 1mm,so to ensure that patients are treated,the IMRT plan to verify and modify the location of the location is very necessary.In the three-dimensional direction,the average precision of the CT simulator and simulation of the positioning machine is controlled within the 2mm.So the two devices can be applied to the location of the IMRT program verification,conducive to improving the use rate and working efficiency of the simulator.The average precision of the CT positioning method is higher than that of the average accuracy of the simulation.Mainly in the Y direction of the difference is obvious.

References:

[1]Hu Yimin,Zhang Hongzhi,Dai Jiangrong,et al.Tumor radiation physics[M].Beijing:Atomic Energy Press,1999:463-470.[胡逸民,张红志,戴建荣,等.肿瘤放射物理学[M].北京:原子能出版社,1999:463-470.]
[2]Zhao Rui,Li Sha,Zeng Limin,et al.Application of nuclertron simulix-HQ simulator in proofreadig isocenter treated[J].Modern Oncology,2011,19(3):550-553.[赵瑞,李莎,曾莉敏,等.核通Simulix-HQ型模拟定位机在靶区治疗中心校位中的应用[J].现代肿瘤医学,2011,19(3):550-553.]
[3]Wang Yonggang,Chen Hong,Liu Yue,et al.Different registration methods on impact of IGRT set-up errors of nasopharyngeal carcinoma treatment[J].Modern Oncolgy,2010,18(4):688-689.[王永刚,陈宏,刘跃,等.不同匹配方式对鼻咽癌IGRT摆位误差的影响[J].现代肿瘤医学,2010,18(4):688-689.]
[4]Chen Jianling,Zhao Yunhui,Chen Chuanben,et al.Analyze positioning repeatability of nasopharyngeal carcinoma by reconstruction image using CT simulator[J].Chin J Med Physics,2012,29(3):3346-3349.[陈建铃,赵云辉,陈传本,等.利用CT模拟机重建图像分析鼻咽癌摆位重复性[J].中国医学物理学杂志,2012,29(3):3346-3349.]
[5]Hurkmans CW,Remeijer P,Lebesque JV,et al.Set-up verification using portal imaging:Review of current clinical pratice[J].Radiather Oncol,2001,58(2):105.
[6]Van Asselen B,Dehnad H,Raaijmakers CP,et al.Implanted gold markers for position verification during irradiation of head-and-neck cansers:a feasibility study[J].Int J Radiat Oncol Biol Phys,2004,59:1011-1017.
[7]Hong TS,Tome WA,Chappell RJ,et al.The impact of daily set-up variations on head-and-neck intensity-modulated radiation therapy[J].Int J Radiat Oncol Biol Phys,2005,61(3):779-788.
[8]Liu Libin,Wu Junxin,Qu Yiyan,et al.Reserch and compare the application of the electronic portal imaging device and cone beam CT in IGRT for the nasopharyngeal carcinoma patients[J].Chin J Med Physics,2012,29(6):3721-3741.[刘利彬,吴君心,瞿宜艳,等.千伏级锥形束CT与兆伏级电子摄野影像系统在鼻咽癌影像引导放疗的对比研究[J].中国医学物理学杂志,2012,29(6):3721-3741.]
[9]Wu Qinhong,Li Xuenan,Liu Chunling,et,al.Evaluation of setup errors for head-and-neck cancer localized with final isocenter marking method via cone beam CT[J].Chin J Radiol Med Prot,2012,32(3):301-303.[吴钦宏,李雪南,刘春玲,等.应用锥形束CT评价最终等中心标记法定位头颈部肿瘤的摆位误差[J].中华放射医学与防护杂志,2012,32(3):301-303.]
[10]Zhang Guojun,Huang Yanping,Shang Jun.Application of CBCT in radiotherapy position of head-and-neck cancer[J].Practical J Cancer,2014,29(10):1334-1335.[张国军,黄艳萍,尚钧.CBCT技术在头颈部肿瘤放射治疗摆位中的应用[J].实用癌症杂志,2014,29(10):1334-1335.]

Memo

Memo:
-
Last Update: 2017-01-26