Influence of different resolutions on pass rate in verification of planar dose distribution of radiotherapy plan

Journal Of Modern Oncology[ISSN:1672-4992/CN:61-1415/R]

Issue:
2022 02
Page:
295-298
Research Field:
Publishing date:

Info

Title:
Influence of different resolutions on pass rate in verification of planar dose distribution of radiotherapy plan
Author(s):
HUANG ShaofuKE QiXU ShaoshengDONG KaibiHUANG Hecheng
Department of Radiation Oncology,the Central Hospital of Shantou,Guangdong Shantou 515031,China.
Keywords:
radiotherapyverification of dose distributionoutput resolutionGamma pass rate
PACS:
R730.55
DOI:
10.3969/j.issn.1672-4992.2022.02.024
Abstract:
Objective:To explore the influence of different resolutions on the pass rate in the verification of planar dose distribution of radiotherapy plan.Methods:40 patients with intensity modulated radiation therapy (IMRT) admitted to our hospital from March 2017 to March 2018 were selected to verify the planar dose distribution of the radiotherapy plan.Different evaluation criteria (3 mm/3%,2 mm/2% and 1 mm/1%) were selected to analyze the change of Gamma pass rate under different evaluation criteria.According to the evaluation criteria,the changes of Gamma throughput under different output resolutions (0.6 mm,1 mm,2 mm,3 mm,4 mm,5 mm,6 mm) were analyzed.Different fields (30 cm×30 cm and 40 cm×40 cm) were selected to analyze the change of Gamma throughput.Results:The Gamma-ray passing rate of 3 mm/3% was higher than 2 mm/2% and 1 mm/1%,and that of 2 mm/2% was higher than 1 mm/1%.The difference was statistically significant (P<0.05).When the output resolution was 3 mm,no matter what evaluation method was used,the pass rate decreased obviously.The pass rate of Gamma-ray at 4 mm,5 mm and 6 mm output resolution was lower than 0.6 mm,1 mm and 2 mm,and the difference was significant (P<0.05).Regardless of the output resolution,the pass rate of 3 mm/3% was always higher than 2 mm/2% and 1 mm/1%,and the pass rate of 2 mm/2% was always higher than 1 mm/1%.The difference was statistically significant (P<0.05).The Gamma-ray pass rate under the field of 30 cm×30 cm was higher than that under 40 cm×40 cm,and the difference was statistically significant (P<0.05).Conclusion:The higher the output resolution is,the lower the pass rate of Gamma is.The pass rate is the highest when 3 mm/3% is used as the evaluation criterion.The pass rate under 30 cm×30 cm field is higher than that under 40 cm×40 cm field.

References:

[1]柏正璐,时飞跃,秦伟,等.绝对剂量的偏移对MatriXX剂量验证结果的影响[J].中国医疗设备,2017,32(10):48-51. BAI ZL,SHI FY,QIN W,et al.Influence of deviation in absolute dose on verification results of MatriXX[J].China Medical Devices,2017,32(10):48-51.
[2]杨玉刚,尚革,许林,等.宫颈癌调强放射治疗的剂量验证分析[J].现代生物医学进展,2014,14(23):4526-4528. YANG YG,SHANG G,XU L,et al.Verification and analysis of the dose of intensity modulation radiated therapy(IMRT) on cervical cancer[J].Progress in Modern Biomedicine,2014,14(23):4526-4528.
[3]郭冉,张艺宝,李明辉,等.调强计划验证通过率的影响因素分析[J].中国医学物理学杂志,2018,35(10):7-11. GUO R,ZHANG YB,LI MH,et al.Factors affecting the passing rate of intensity-modulated radiotherapy plans[J].Chinese Journal of Medical Physics,2018,35(10):7-11.
[4]陈利军,谷晓华,杨留勤.位置移动和剂量调节对调强放射治疗剂量验证通过率的影响[J].医疗卫生装备,2017,38(6):115-117. CHEN LJ,GU XH,YANG LQ.Effects of position shift and dose adjustment on pass rate of IMRT dose verification[J].Chinese Medical Equipment Journal,2017,38(6):115-117.
[5]刘礼东,杨振,邱小平,等.食管癌调强计划三维剂量验证中γ通过率最佳阈值的研究[J].中华放射医学与防护杂志,2018,38(4):297-301. LIU LD,YANG Z,QIU XP,et al.The optimal gamma passing rate thresholds of IMRT dosimetric verification in the treatment of esophageal cancer[J].Chinese Journal of Radiological Medicne and Protection,2018,38(4):297-301.
[6] 王庆章,段天宇,刘建超,等.Monaco与Pinnacle治疗计划系统在食管癌放疗剂量验证上的比较[J].中国医学物理学杂志,2017,34(2):166-169. WANG QZ,DUAN TY,LIU JC,et al.Comparison of Monaco and Pinnacle treatment planning systems on the dosimetric verification of radiotherapy for esophageal cancer[J].Chinese Journal of Medical Physics,2017,34(2):166-169.
[7] 李娜,冯革,张旭,等.容积旋转调强计划计算分辨率对COMPASS验证通过率的影响[J].中华放射医学与防护杂志,2018,38(10):761-766. LI N,FENG G,ZHANG X,et al.The impact of dose grid resolution during VMAT planning on COMPASS pass rate[J].Chinese Journal of Radiological Medicne and Protection,2018,38(10):761-766.
[8]LEE R,KIM K,CHO S,et al.Statistical process control analysis for patient quality assurance of intensity modulated radiation therapy[J].Journal of the Korean Physical Society,2017,71(10):717-721.
[9]NJEH C,SALMON H,SCHILLER C.The impact of dose rate on the accuracy of step-and-shoot intensity-modulated radiation therapy quality assurance using varian 2300CD[J].Journal of Medical Physics,2017,42(4):206-212.
[10]WOON WA,RAVINDRAN PB,EKAYANAKE P,et al.A study on the effect of detector resolution on gamma index passing rate for VMAT and IMRT QA[J].Journal of Applied Clinical Medical Physics,2018,19(2):230-248.
[11]RADHAKRISHNAN S,CHANDRASEKARAN A,SARMA Y,et al.Dosimetric comparison between single and dual arc-volumetric modulated arc radiotherapy and intensity modulated radiotherapy for nasopharyngeal carcinoma using a simultaneous integrated boost technique[J].Asian Pacific Journal of Cancer Prevention,2017,18(5):1395-1402.
[12]时飞跃,柏正璐,唐小羽,等.IGRT治疗床模型对食管癌调强计划ArcCHECK剂量验证的影响[J].中国医疗设备,2018,33(10):91-93. SHI FY,BAI ZL,TANG XY,et al.Influence of treatment couch model on dose verification of intensity-modulated radiotherapy plans of esophageal cancer with ArcCHECK[J].China Medical Devices,2018,33(10):91-93.
[13]DU FL,ZHAO K,SHI GD,et al.Dose verification of constant dose rate-volumetric modulated arc therapy plan on Varian clinical 23EX linear accelerator through MapCheck phantom[J].Chinese Journal of Medical Physics,2017,34(8):776-782.
[14]JIA F,ZHANG YB,WANG MJ,et al.Effects of different algorithms on quality assurance of compass-based radiotherapy plan[J].Chinese Journal of Medical Physics,2017,34(4):330-334.
[15]周晖,许赪,沈文同,等.左乳癌保乳术后调强放疗计划三维剂量验证[J].中国医学物理学杂志,2017,34(12):1216-1219. ZHOU H,XU C,SHEN WT,et al.Three-dimensional dose verification of intensity-modulated radiotherapy after left breast-conserving surgery[J].Chinese Journal of Medical Physics,2017,34(12):1216-1219.

Memo

Memo:
广东省汕头市科技计划医疗卫生类别项目(编号:181204064010552)
Last Update: 1900-01-01