Clinical application of thermoplastic membrane plus balloon combined with OPS technique in radiotherapy of liver cancer

Journal Of Modern Oncology[ISSN:1672-4992/CN:61-1415/R]

Issue:
2020 03
Page:
459-462
Research Field:
Publishing date:

Info

Title:
Clinical application of thermoplastic membrane plus balloon combined with OPS technique in radiotherapy of liver cancer
Author(s):
Liu YuGe BinbinWang JunhuiZhao YongliangLiu HaitaoChu Kaiyue
Radiology Department,Nantong Tumor Hospital,Nantong University Affiliated Tumor Hospital,Jiangsu Nantong 226361,China.
Keywords:
simple abdominal pressureoptical positioning systempatient setup errorliver cancer
PACS:
R735.7
DOI:
10.3969/j.issn.1672-4992.2020.03.024
Abstract:
Objective:Simple abdominal pressure (SAP) combined with optical positioning system was designed to study the effect of placement on liver cancer radiotherapy and its clinical application value in precise radiotherapy of thoracic and abdominal tumors.Methods:This study was divided into two comparison groups according to the placement technique of liver cancer patients.Each group selected 15 patients,which was the control group with free breathing routine positioning and the experimental group of SAP combined with OPS technique to control the respiratory amplitude localization.Firstly,the tumor motility of the two groups of patients under the simulator was observed,and then the design plan of the target area was delineated.Before the treatment,the cone beam computed tomography (CBCT) was performed to verify the tumor motility and treatment.The linearity error was analysed at the time of X (left and right direction),Y (head and foot direction),and Z (face and back direction).Results:The tumor motility of the experimental group and the control group in the X,Y,and Z directions under the simulator were (2.461 5±0.660 2,3.923 1±1.187 5)mm,(5.692 3±1.548 4,19.076 9±4.499 3)mm,(3.230 8±1.165 8,8.692 3±2.982 9)mm,and the corresponding P values in each direction were P=0.001,P<0.001,P<0.001.The linear positioning errors of the experimental group and the control group in the X,Y,and Z directions were respectively (2.175 3±0.464 8,4.940 0±0.573 3)mm,(1.435 3±0.304 9,6.408 0±0.555 0)mm,(2.288 7±0.591 4,4.967 3±0.353 9)mm,and the corresponding P values in all directions are less than 0.001.Conclusion:Compared with the conventional positioning technology,the designed SAP combined with OPS technology can significantly reduce the tumor respiratory motility in the treatment of liver cancer patients and improve the positioning accuracy during treatment.And the treatment accuracy of chest and abdomen tumors has clinical application value.

References:

[1]LI Gong.Research progress in radiotherapy for liver cancer[J].Journal of Oncology,2010,16(7):516-519.[黎功.肝癌放疗研究进展[J].肿瘤学杂志,2010,16(7):516-519.]
[2]CHEN Mingming,CHEN Ying,CHEN Xiangning,et al.Development of medical accelerator positioning control system based on ARM[J].Chinese Journal of Medical Physics,2015,32(6):790-793.[陈明明,陈颖,陈相宁,等.基于ARM的医用加速器摆位控制系统研制[J].中国医学物理学杂志,2015,32(6):790-793.]
[3]Stroom JC,Heijmen BJ.Geometrical uncertainties,radiotherapy planning margins,and the ICRU-62 report[J].Radiother Oncol,2002,64(1):75-83.
[4]Sarrut D,Boldea V,Ayadi M,et al.Nonrigid registration method to assess reproducibility of breath-holding with ABC in lung cancer[J].Int J Radiat Oncol Biol Phys,2005,61(2):594-607.
[5]Cheung PC,Sixel KE,Tirona R,et al.Reproducibility of lung tumor position and reduction of lung mass with in the planning target volume using active breathing control (ABC)[J].Int J Radiat Oncol Biol Phys,2003,57(5):1437-1442.
[6]Dawson LA,Brock KK,Kazanjian S,et al.The reproducibility of organ position using active breathing control(ABC) during-liver radiotherapy[J].Int J Radiat Oncol Biol Phys,2001,51(5):1410-1421.
[7]ZHAO Jiandong,XU Zhiyong,HU Weigang,et al.Feasibility of applying active breathing control (ABC)for 3-dimensional conformal radiation therapy (3DCRT)in hepatocellular carcinoma (HCC)and dosimetric study[J].Oncology Progress,2006,4(4):319-326.[赵建东,徐志勇,胡伟刚,等.主动呼吸控制技术用于原发性肝癌放疗的可行性及剂量学研究[J].癌症进展杂志,2006,4(4):319-326.]
[8]ZHANG Li.Application of active breathing control and dosimetric analysis for non-small cell lung cancer[D].Beijing:Tsinghua University,2009.[张黎.主动呼吸控制技术在肺癌放射治疗中的应用及剂量学研究[D].北京:清华大学,2009.]
[9]CHEN Xinde.Application of respiratory gating in radiation therapy of primary liver cancer[J].Practical Clinical Journal of Integrated Traditional Chinese and Western Medicine,2014,14(10):71-72.[陈心得.呼吸门控在原发性肝癌放射治疗中的应用[J].实用中西医结合临床,2014,14(10):71-72.]
[10]Herfarth KK,Debus J,Lohr F,et al.Extracranial stereotactic radiation therapy:set-up accuracy of patients treated for liver metastases[J].Int J Radiat Oncol Biol Phys,2000,46(2):329-335.
[11]Negoro Y,Nagata Y,Aoki T,et al.The effectiveness of an immobilization device in conformal radiotherapy for lung tumor:Reduction of respiratory tumor movement and evaluation of the daily setup accuracy[J].Int J Radiat Oncol Biol Phys,2001,50(15):889-898.
[12]ZHAO Hui,ZHAO Wenhui,SHI Fang,et al.Improvement on target positioning precision in radiation therapy IGRT[J].Modern Oncology,2013,21(06):1351-1353.[赵辉,赵文晖,师芳,等.提高放射治疗摆位精度[J].现代肿瘤医学,2013,21(06):1351-1353.]
[13]Lei Chai,Da Chen,Xiao-Bin Tang,et al.Radiotherapy treatment based on binocular location:Real-time monitoring system of patient setup errors[J].Journal of Medical Imaging and Health Informatics,2017(7):897-902.
[14]QIU Teng,ZHAO Di,SONG Wei,et al.Application of infrared positioning system in the placement and verification of radiotherapy for prostate cancer[J].Chin Cini Oncol,2018,23(3):255-259.[邱腾,赵迪,宋威,等.红外定位系统在前列腺癌放射治疗的摆位和验证中的应用[J].临床肿瘤学杂志,2018,23(3):255-259.].
[15]YANG Xiaomei,CHU Kaiyue,JIN Jianhua,et al.Application of chest positioning reference line combined with infrared positioning system in intensity-modulated radiotherapy for nasopharyngeal carcinoma[J].Chinese Journal of Medical Physics,2017,34(2):147-150.[杨晓梅,储开岳,金建华,等.胸部定位参考线配合红外定位系统在鼻咽癌放疗中应用[J].中国医学物理学杂志,2017,34(2):147-150.]

Memo

Memo:
江苏省重点研发计划(编号:BE2017679)
Last Update: 1900-01-01